NOMEX
Operating Manual - Equipment Description
D909.131.00/02 en
17
Operating Manual
1 Equipment Description
1.1
Features
– multi-channel semiconductor detector
– measurement of dose, dose rate, dose per
pulse, pulses, pulse frequency, exposure time,
tube voltage, total filtration, and half value
layer as well as of the tube voltage and dose
rate waveforms
– fully automatic dose and tube voltage ranging
– fully automatic total filtration compensation
– measurement of very heavily filtered radiation
qualities
– measurement of the half value layer in all appli-
cations
– adjustable sampling with an appropriately
sized log memory for several seconds
– adjustable timer functions
– message if the accuracy of the measured
value is inadequate
– auto-start/auto-stop mode with auto-reset
– USB interface
– powered via USB interface
– NOMEX software for the display and evalua-
tion of the measured values
– NOMEX software for equipment setup
– selectable display language for the NOMEX
software
CAUTION
Improper handling.
Equipment Damage!
The NOMEX multimeter is a sensitive measur-
ing device that requires regular inspection and
calibration.
– Ensure that the NOMEX multimeter is ser-
viced at regular intervals as outlined in
section
.
– Have the NOMEX multimeter calibrated
periodically by an authorized laboratory
(also refer to section
Legally Calibrated NOMEX Multimeters
The calibration authority has placed a calibration
mark on the back of the device. Damaging or
removing the calibration mark automatically inval-
idates the calibration.