webline

 

 

Regularly Maintenance

 

 

D776.133.0/1

3 Regularly 

Maintenance 

3.1  Checkups (Function tests) 

The following checkups have to be executed regu-

larly according to the recommendation or the na-

tionally valid laws / rules.  

 

Test 
 

Interval Recommendation 

Comment 

Visual inspection of the device 
and the cables  
 

daily and after  
breaks (over 
several hours) 
 

strongly recom-
mended 

 

Current zeroing 

daily and after  
breaks (over 
several hours) 
 

strongly recom-
mended 

repeat after each changing of the 
detector  

Detector test 

daily and after  
breaks (over 
several hours), 
after changing 
the detector 
 

recommended 

HV check at detector 

Check desiccator 
 

weekly recommended 

 

Safety inspections 
 

annually 

acc. law 

refer to national laws / rules 
 

Inspections of measurement 
technique  
 

2 years 

acc. law 

refer to national laws / rules 
 

 
To execute the tests correctly, please observe the 

instructions given in the corresponding sections of 

the user manual of the UNIDOS

webline

.  

Safety inspections must only be handled by per-

sons who are trained in the use of such equipment 

and are capable of applying it properly. The opera-

tor must be trained in the use of the device. 

The UNIDOS

webline

 must be inspected at regular 

intervals according to national regulations at PTW-

Freiburg or a qualified calibration laboratory.  

Safety inspections and inspections of measurement 

technique must also be performed after repairs or 

after each influence which may have changed the 

behavior of the device. 

National valid laws or rules may regulate shorter 

intervals. 

 

 

 

Summary of Contents for UNIDOS webline

Page 1: ...Service Manual UNIDOSwebline D776 133 0 1 2006 06 Se Sa...

Page 2: ...rts List 3 3 Regularly Maintenance 4 3 1 Checkups Function tests 4 3 2 Constancy Test 5 3 3 Maintenance 5 3 4 Further Tests 6 3 5 System Information 6 4 Troubleshooting 7 5 Functional units 7 5 1 Fron...

Page 3: ...rfaces described in this manual must fulfill the re quirements of the IEC standard 60950 PTW FREIBURG Physikalisch Technische Werkst tten Dr Pychlau GmbH L rracher Str 7 79115 FREIBURG GERMANY Phone 4...

Page 4: ...Safety inspections annually acc law refer to national laws rules Inspections of measurement technique 2 years acc law refer to national laws rules To execute the tests correctly please observe the in...

Page 5: ...corresponding sections of the user manual of the UNIDOSwebline The test results have to be recorded from the first start of operation in order to be able to make state ments about a change Addionally...

Page 6: ...mail on a service e mail at unidos webline9 ptw de UNIDOS webline SERVICE 2006 02 01 12 04 09 System information Device name UNIDOS webline Type number T10021 Serial number 5 Hardware revision A02 Fir...

Page 7: ...en tire enclosure of the measuring module is on the potential of the high voltage In order to switch off the device safely it must be turned off and separated from the power line Moreover all connect...

Page 8: ...lection HV polarity Electrical Isolation Amplifier HV to detector HV measurement Figure 2 Measuring module The measuring module is designed as an inde pendent unit and can be exchanged with another on...

Page 9: ...rd contains the individual subunits for generation and monitoring of the supply voltages as well as for charging the rechargeable batteries It fulfills connection functions beside that Its name origin...

Page 10: ...rmware Restore of the device settings as well as the detector data base If a firmware update from PTW Freiburg is released for the UNIDOSwebline the user can update the firm ware version on the newest...

Reviews: