background image

UNIDOS

webline

 

Regularly Maintenance

 

 

D776.133.0/1 

5

3.2 Constancy Test 

The following tests should be executed regularly. 

They make it possible to recognize slow alterations 

of the characteristics of device or detector. 

 

Test 
 

Interval 

Accessories
 

Comment 

Detector test 
 

weekly 

HV check at detector; only range Low 

Device test 
 

weekly 

checks device only; only range Low 

Long term stability  
 

weekly 

UNITEST 

check HV and all 3 measuring ranges; 
no complete function test 
 

Long term stability 
 

monthly 

check device 

no complete function test 

 
To execute the tests correctly, please observe the 

instructions given in the corresponding sections of 

the user manual of the UNIDOS

webline

The test results have to be recorded from the first 

start of operation in order to be able to make state-

ments about a change. Addionally for detector 

tests: the record has to be separate for each detec-

tor.  

The checkup of the device can be performed most 

extensive with an UNITEST (refer to the user man-

ual of the UNITEST).   

A complete function test can take place however 

only through a inspection of measurement tech-

nique. 

3.3 Maintenance 

 
Description 
 

Interval Comment 

Cl care 
 

if required 

 
 

Rechargeable battery 
- handling – 
 

max. 3 month 

discharge and charge NiMH batteries 

Changing fuses 
 

if required 

 

 
To execute the tests correctly, please observe the 

instructions given in the corresponding sections of 

the user manual of the UNIDOS

webline

 

Summary of Contents for UNIDOS webline

Page 1: ...Service Manual UNIDOSwebline D776 133 0 1 2006 06 Se Sa...

Page 2: ...rts List 3 3 Regularly Maintenance 4 3 1 Checkups Function tests 4 3 2 Constancy Test 5 3 3 Maintenance 5 3 4 Further Tests 6 3 5 System Information 6 4 Troubleshooting 7 5 Functional units 7 5 1 Fron...

Page 3: ...rfaces described in this manual must fulfill the re quirements of the IEC standard 60950 PTW FREIBURG Physikalisch Technische Werkst tten Dr Pychlau GmbH L rracher Str 7 79115 FREIBURG GERMANY Phone 4...

Page 4: ...Safety inspections annually acc law refer to national laws rules Inspections of measurement technique 2 years acc law refer to national laws rules To execute the tests correctly please observe the in...

Page 5: ...corresponding sections of the user manual of the UNIDOSwebline The test results have to be recorded from the first start of operation in order to be able to make state ments about a change Addionally...

Page 6: ...mail on a service e mail at unidos webline9 ptw de UNIDOS webline SERVICE 2006 02 01 12 04 09 System information Device name UNIDOS webline Type number T10021 Serial number 5 Hardware revision A02 Fir...

Page 7: ...en tire enclosure of the measuring module is on the potential of the high voltage In order to switch off the device safely it must be turned off and separated from the power line Moreover all connect...

Page 8: ...lection HV polarity Electrical Isolation Amplifier HV to detector HV measurement Figure 2 Measuring module The measuring module is designed as an inde pendent unit and can be exchanged with another on...

Page 9: ...rd contains the individual subunits for generation and monitoring of the supply voltages as well as for charging the rechargeable batteries It fulfills connection functions beside that Its name origin...

Page 10: ...rmware Restore of the device settings as well as the detector data base If a firmware update from PTW Freiburg is released for the UNIDOSwebline the user can update the firm ware version on the newest...

Reviews: