background image

UNIDOS

webline

  

Functional units

 

 

D776.133.0/1 

9

5.3 Main Board 

 

Mains 

Connector & 

Filter

Rechargeable 

batteries

Charge 

control

Power 
supply 

units

Voltages 

measurement

Relay

Wide 

range

AC / DC

Supply voltages

Host I/O (SERIAL)

Ethernet

TRIP Output

Electrical Isolation

Rear panel

 

Figure 3: Main board 

The main board contains the individual subunits for 

generation and monitoring of the supply voltages as 

well as for charging the rechargeable batteries. It 

fulfills connection functions beside that. Its name 

originates from its physical size and implies in no 

way, that it houses the main CPU. 

The calibration factors for measuring of the supply 

voltage are stored in the main board. 

The results of the voltage monitoring can be read 

out via service e-mail. It is not necessary to execute 

measurements inside the device. 

 

 

Summary of Contents for UNIDOS webline

Page 1: ...Service Manual UNIDOSwebline D776 133 0 1 2006 06 Se Sa...

Page 2: ...rts List 3 3 Regularly Maintenance 4 3 1 Checkups Function tests 4 3 2 Constancy Test 5 3 3 Maintenance 5 3 4 Further Tests 6 3 5 System Information 6 4 Troubleshooting 7 5 Functional units 7 5 1 Fron...

Page 3: ...rfaces described in this manual must fulfill the re quirements of the IEC standard 60950 PTW FREIBURG Physikalisch Technische Werkst tten Dr Pychlau GmbH L rracher Str 7 79115 FREIBURG GERMANY Phone 4...

Page 4: ...Safety inspections annually acc law refer to national laws rules Inspections of measurement technique 2 years acc law refer to national laws rules To execute the tests correctly please observe the in...

Page 5: ...corresponding sections of the user manual of the UNIDOSwebline The test results have to be recorded from the first start of operation in order to be able to make state ments about a change Addionally...

Page 6: ...mail on a service e mail at unidos webline9 ptw de UNIDOS webline SERVICE 2006 02 01 12 04 09 System information Device name UNIDOS webline Type number T10021 Serial number 5 Hardware revision A02 Fir...

Page 7: ...en tire enclosure of the measuring module is on the potential of the high voltage In order to switch off the device safely it must be turned off and separated from the power line Moreover all connect...

Page 8: ...lection HV polarity Electrical Isolation Amplifier HV to detector HV measurement Figure 2 Measuring module The measuring module is designed as an inde pendent unit and can be exchanged with another on...

Page 9: ...rd contains the individual subunits for generation and monitoring of the supply voltages as well as for charging the rechargeable batteries It fulfills connection functions beside that Its name origin...

Page 10: ...rmware Restore of the device settings as well as the detector data base If a firmware update from PTW Freiburg is released for the UNIDOSwebline the user can update the firm ware version on the newest...