Please Read Before Installation
The following will damage the Air Check Oxygen monitor.
1. The CO
2
monitor requires
24 VDC regulated power
.
Please Do Not connect
the monitor to any voltage that exceeds
24 Volts DC, or
Any AC Voltage.
2. Do not power the carbon dioxide sensor unplugged from the main PC
board.
Do Not Connect
the CO
2
sensor to the PC board while the
monitor is powered. This Will Damage the sensor.
3. The CO
2
cell range is matched to the electronics.
Never exchange
the
electronics with a CO
2
sensor from a different monitor.
4. When calibrating or challenging the CO
2
monitor,
a. Do not expose the monitor to flow rates that exceed ½ liter per
minute, (500 cc per minute) flow.
b. Expose the monitor to span gas blends that consist of
Oxygen, Nitrogen or Carbon Dioxide only.
Do Not expose
the monitor to any combustible gas,
i.e. Methane, Hydrogen,
etc.
5. Do not expose the CO
2
monitor to silicone compounds. They can
cause a loss of sensitivity.
6. Do not expose the monitor to high flow air or install it directly in front of
fans.
7. When using the CO
2
monitor, do not expose the sensor directly to a
water stream. In areas requiring wash downs, cover and protect the
monitor and power supply. Contact PureAire for details on a waterproof
enclosure.
8. Please refer to section 6.3 of this manual regarding the CO
2
sensor.
YOU MUST
choose a calibration method before use.