6: Maintenance & Sensor Verification
Only qualified personnel should perform maintenance and sensor verification
6.1.1
Sensor Cell Recharge
1.
Remove the sensor cell from the Universal Gas Detector or duct.
2.
Remove the box nut, membrane retainer ring, membrane, O-ring and silicon sheet, (if provided)
3.
Do not remove the cell from the mounting ring
NOTE:
The O-rings can be either black or white. Refer to section 6.1.2 for sensor spare parts
Box Nut
Membrane Retainer Ring
Membrane
O-ring
Sensing Electrode
Silicon Sheet
4.
Place the sensor over the beaker as shown and pour the old electrolyte into the beaker.
Pressure
Compensation
Screw
Beaker for
Old Electrolyte
NOTE: The sensor
has a breather plug
on the side of the
sensor.
It is covered with a
white membrane, the
Miliseal, p/n SC2009
DO NOT REMOVE
THE MILISEAL OR
BREATHER PLUG
NOTE:
Some
sensor cells have a
clear silicone
diffuser installed
inside the
Membrane
Retaining Ring.
DO NOT
REMOVE IT
NOTE: Some
sensors have a white
miliseal attached to
the box nut.
DO NOT
REMOVE IT