PureAire Monitoring Systems, Inc.
45
Sensors
Gas
Range
Electrolyte
Membrane
White or
Black O-ring
Flat O-Ring
GS-4060
AsH
3
1 ppm
EL-4070-7A
M-4000G
White
Yes
GS-4060
PH
3
1 ppm
EL-4070-7A
M-4000G
White
Yes
GS-4060
SiH
4
15 ppm
EL-4070-7A
M-4000G
White
Yes
GS-4060
H2Se
1 ppm
EL-4070-7A
M-4000G
White
Yes
7
8
9
6.2 Sensor Calibration Procedure
CAUTION: Be sure to observe all safety guidelines when using span gases.
NOTE:
If the instrument is connected to a controller, set the controller to the
standby mode to avoid accidental alarms. The Universal Gas Detector’s
internal alarm LED’s and relays are always live. During a calibration
test if levels exceed preset alarm thresholds you will also be verifying the
internal alarms and any remote horns, strobes or fans that are connected.
6.2.1 Sensor Gas Calibration
The Universal Gas Detector requires periodic calibration with the appropriate
calibration gas. A calibration test is a quantitative test of the Universal Gas
Detector to a known gas concentration certified by a gas supplier.
PureAire recommends the use of specialty gas suppliers for obtaining calibration
gas standards. All span gases must be a blend of the target gas blended in air.
NOTE:
Span gas must be blended with air. Do not blend with nitrogen
The target gas concentration should be in the middle of the dynamic range of the
detector, i.e. (for a 0-10ppm range use 5ppm gas).
In the absence of a known span gas, a qualitative bump gas test may be
performed using commercially available products, i.e. Clorox for testing
chlorine, bromine and fluorine detectors or ammonia for testing NH
3
detectors.
Consult PureAire for more information on test gases.
Calibration should be performed whenever:
•
The membrane or electrolyte is replaced
•
The entire sensor cell is replaced
•
Six months has passed without membrane, electrolyte, or sensor replacement
NOTE
: for higher accuracy more frequent dynamic gas calibration is required.
*Black O-ring part number
(
1516-12W)
, White O-ring part number (
1516-12B
), Flat O-ring Silcon sheet part
no.
(SC2000)*
**For pricing and part number questions please consult your sales rep. If part number is unavailable or sensor is