PureAire Monitoring Systems, Inc.
46
6.2.2 Sensor Calibration Equipment
It is recommended to purchase calibration gas standards directly from your
specialty gas provider. Calibration span gas standards can be found from the
following suppliers:
Air Liquide Specialty Gas; (
alspecialtygases.com
)
Scott Specialty Gas; (
alspecialtygases.com
)
CalGas; (
alspecialtygases.com
)
AirGas; (
www.airgas.com
)
The following equipment is required to facilitate gas calibration:
Part Number
Description
Quantity
Provided by Gas Supplier
104 liter cylinder span gas, balance air
1
CZF7R000255 (CalGas p/n)
Regulator, Model 715, (500 cc/min flow)
1
83505 (PureAire p/n)
Calibration Cap
1
90054
Span gas tubing Assy
1
6.2.3 Sensor Calibration Procedure
1.
If the instrument is connected to a controller, set the controller to Standby
mode to avoid accidental alarms.
2.
Remove the smoked front transmitter cover.
3.
Enter the password to permit access to the menus
4.
Push the joystick to the Right once to go into each adjustment menu. You’ll
see the first menu, Set 4-20mA loop.
5.
Push the joystick Up twice until the “Sensor Adjustment” menu is displayed.
6.
Push the joystick Right to display the “Set Sensor Span” menu
...PassCode OK……
0.0 ppm
.Set 4-20mA loop…
0.0 ppm
.Sensor Adjustment..
0.0 ppm
..Set Sensor Span…
0.0 ppm