P8400W Air-dryer User’s Guide
Page 66 of 110
P017585 – Rev. A
9.2 Measuring Compressor Amp Draw
With the Compressor running:
9.2.1
Remove Front Panel (see section 8.10.2 )
9.2.2
Locate the hot lead wire
going to the Compressor you
will be measuring:
System 1 =
Wire #1
System 2 =
Wire #8
9.2.3
Use an Amp Meter to
measure the running amps.
With the compressor running, the
running amps should measure
4.0 amps or below
.
9.2.4
Reinstall Front Panel.
If the compressor measures over 4.0 running amps, see section 13.15 for
troubleshooting information.
WARNING!
Internal surfaces may be hot
. Use care when coming into contact
with internal components as there is a potential for some these
components to become hot when in operation or standby.