ALTEC AIR, LLC
P1500W Series Air Dryers User’s Guide
Page 70 of 100
P011787 – Rev. S
9.14.6
Push knob in to lock (or tighten the retaining nut – LP Models).
9.14.7
Press the
RESET
Button
.
9.14.8
Close Front Panel.
If you are unable to create a High Outlet Pressure Alarm as described, see
section 13.6 for troubleshooting information.
9.15 Testing Low Outlet Pressure Alarm
9.15.1
Make a note of the current
Outlet Pressure (
OUTP
)
reading.
9.15.2
Open Front Panel
(section 8.6 ).
9.15.3
Pull the Outlet Pressure
Regulator knob out (or
loosen the retaining nut – LP
Models).
9.15.4
Turn knob counter-
clockwise until Outlet
Pressure (
OUTP
) reading drops
below:
6.5 PSI
for P1500W & P1502W models
0.30 PSI
for P1500WLP& P1502WLP models
25.0 PSI
for P1500WHP & P1502WHP models
After one (1) minute, the Low Pressure Alarm should appear on the display.