ALTEC AIR, LLC
P1500W Series Air Dryers User’s Guide
Page 69 of 100
P011787 – Rev. S
9.14 Testing High Outlet Pressure Alarm
9.14.1
Make a note of the current
Outlet Pressure (
OUTP
)
reading.
9.14.2
Open Front Panel (section
9.14.3
Pull the Outlet Pressure
Regulator knob out (or loosen
the retaining nut – LP
Models).
9.14.4
Turn knob clockwise until
Outlet Pressure (
OUTP
)
reading climbs
over:
12.0 PSI
for P1500W & P1502W models
7.50 PSI
for P1500WLP& P1502WLP models
35.0 PSI
for P1500WHP & P1502WHP models
After one (1) minute, the High Pressure Alarm should appear on the display.
9.14.5
Turn Outlet Pressure Regulator knob counter-clockwise until Outlet
Pressure (
OUTP
) reading
lowers to the reading
recorded in step 9.14.1