PUREGAS, LLC
P550W2 Series Air Dryers User’s Guide
Page 70 of 99
P012226 – Rev. F
Allow the Humidity reading to rise
over 10.0%.
9.12.5
After three (3) minutes,
verify that a Humidity Alarm
appears and the Dryer goes
into
SHUTDOWN
mode.
9.12.6
Replace the Humidity
Sensor into the Humidity
Block.
9.12.7
Press the
RESET Button
to clear the Humidity alarm.
If you are unable to create a Humidity / Shutdown alarm as described, see section
13.10 for troubleshooting information.