9.13 Testing Humidity Alarm and System Shutdown
NOTE:
For this test, make sure the air dryer is operating in
BOTH
Cycle Mode
(see section 8.4.2 for details on changing Cycle Mode)
.
Test one System at a time (
System* = System 1 or System 2).
9.13.1
Remove Front Panel (see section 8.7.2 )
9.13.2
Turn the System* Circuit
Breaker
OFF
.
9.13.3
When the System* Screen
(8.4.5.3 ) appears on the
display, press the
HOLD
Button
on the Front Panel to
freeze that screen.
9.13.4
Verify the System*
pressure is zero (0).