PUREGAS, LLC
P6500W Air Dryer User’s Guide
Page 58 of 88
P010452 – Rev. Q
9.14.6
Turn Outlet Pressure
Regulator knob counter-
clockwise until Outlet
Pressure (
OUTP
) reading
lowers to the reading recorded
in step 9.14.3
9.14.7
Push knob in to lock.
9.14.8
Press the
RESET Button
to clear the alarm.
9.14.9
Reinstall Front Panel.
If you are unable to create a High Outlet Pressure Alarm as described, see
section 13.6 for troubleshooting information.
9.15 Testing Low Outlet Pressure Alarm
9.15.1
Remove Front Panel (see section 8.7.2 )
9.15.2
When the Unit Screen
(8.4.5.1 ) appears on the
display, press the
HOLD
Button
on the Front Panel to
freeze that screen.
9.15.3
Make a note of the current
Outlet Pressure (
OUTP
)
reading.