PUREGAS, LLC
P6500W Air Dryer User’s Guide
Page 59 of 88
P010452 – Rev. Q
With Compressor(s) running:
9.15.4
Pull the Outlet Pressure
Regulator knob out.
9.15.5
Turn knob counter-
clockwise until Outlet
Pressure (
OUTP
) reading
drops
below 6.5 PSI
.
After one (1) minute, the Low
Pressure Alarm should appear on the display.
9.15.6
Turn Outlet Pressure
Regulator knob clockwise
until Outlet Pressure (
OUTP
)
reading raises to the reading
recorded in step 9.15.3
9.15.7
Push knob in to lock.
9.15.8
Press the
RESET Button
to clear the alarm.
9.15.9
Reinstall Front Panel.
If you are unable to create a Low Outlet Pressure Alarm as described, see section
13.8 for troubleshooting information.