How to operate the
KnightStar 330
3
Y-500009-00 Rev. J (6/06)
KnightStar 330
User’s Manual
3-11
To run the
KnightStar 330
using the external battery:
1. Turn the
KnightStar 330
off and disconnect the AC
power cord from the back of the device.
2. Connect one end of the appropriate cable to the
battery or automobile cigarette lighter/power port and
the other end to the external battery connector at the
back of the
KnightStar 330
(refer to Figure 1-2 and Table
3. Turn the
KnightStar 330
on.
To switch back to AC power:
1. Turn the
KnightStar 330
off and disconnect the battery
cable from the back of the device.
2. Connect the AC power cord to the
KnightStar 330
and
plug into a wall outlet.
3. Turn the
KnightStar 330
on.
3.5 How to clean the
KnightStar 330
It is important to clean all of the equipment regularly. Do
not use any cleaning method other than described here.
3.5.1 How to clean the breathing circuit
Before you use the breathing circuit for the first time (and
at least three times per week thereafter):
1. Plug the ends of the patient pressure line (small tube
running along the length of the circuit) with the plugs
that came with the circuit.
2. Wash the circuit with warm, soapy, distilled water.
Connect the
KnightStar 330
to only one power
source at a time; external battery or AC power.
Damage to the device can result if connected to
both power sources at the same time.