3
How to operate the
KnightStar 330
3-12
KnightStar 330
User’s Manual
Y-500009-00 Rev. J (6/06)
3. Rinse thoroughly with distilled water to remove soap
residue.
4. Disinfect according to the instructions for use supplied
with your breathing circuit.
5. Allow the breathing circuit to dry. Remove the plugs
from the patient pressure line before use.
3.5.2 How to clean the exterior surfaces
Clean the surfaces of the
KnightStar 330
by wiping with a
cloth dampened with warm, soapy water, then wiping dry
.
To disinfect the exterior surfaces, you may wipe them with
a cloth dampened with:
•
Isopropyl alcohol (70% solution), or
•
A mixture of 1 part household bleach to 10 parts
distilled water (10% solution)
3.5.3 How to clean the inlet air filter
Inspect the inlet filter often by removing the inlet baffle
(removable plastic cover on the back of the unit). Clean
the filter once a week or more often in dusty
environments.
1. Remove the baffle from the back of the unit
(Figure 3-5) by pulling outward on the middle of the
baffle while pushing one end toward the center.
Always unplug the
KnightStar 330
from all
electrical power sources before cleaning. Do
not let water drip into any opening on the
unit.
Replace the inlet filter if torn or damaged in any
way.