2.
The pictures on pages 9 and 10 show the available Wraps. Pictures
demonstrating the Squid Go being used in different treatment
areas are found on pages 11-22.
The Squid Go must be used
over clothing or over the Barrier Cloth.
Place the Wrap over the
clothing or Barrier Cloth with the Cold Gel Pack facing the skin and
the Tube Connector towards the ground.
3.
Secure the Wrap in place using the Velcro.
4.
Connect both ends of the tubing to the Connectors on the Wrap
and Pump control unit by pressing the metal clip on the Tube side
of the Connector.
Medical clip on Tube side
of the Connector
INITIATING TREATMENT:
1.
Turn the control unit
ON
by pressing and holding the “
POWER
”
button for half a second.
2.
Select an operation mode by pressing the “
MODE
” selection
button . The selected mode will appear on the display as a
number from 1 to 4 (1 for low pressure and 4 for high pressure).
Always start using in mode 1. Use modes 2, 3, 4 to increase
pressure until a comfortable pressure is achieved. The mode may
be changed during use by pressing the “
START
” button for
half a second. When the screen begins to blink, press the “
MODE
”
selection button to the level desired. Once the level desired
is flashing on the screen, press the “
START
” button again to
continue the treatment.
3.
Press the “
START
” button for half a second to initiate
treatment. The device will intermittently compress the Cold Gel
Pack to the skin for 15 min.
4.
When 15 minutes are over and the timer on the Pump control unit
is showing (0), remove the Wrap from the anatomical location. The
Squid Go Pump control unit will turn off automatically after less
than 5 seconds.