PTC System Controls and Diagnostics
CP15 User Manual CP15_UM_221017 Page 28 of 33
14
Fault-finding
Symptom
Possible Cause
Confirmation
Solution
No output pulses seen
No signal from detector.
Connect known good signal
source.
Correct any detector or
radiation source issue
Very low pulse rate
Set up oscilloscope to capture
individual events.
Increase radiation flux if
possible.
Insufficient gain
Look for pulses with
oscilloscope.
Increase gain setting.
Increase detector gain if
possible (photomultiplier
voltage).
Reduce threshold of pulse
counting discriminator circuits
if possible.
Wrong output pulse polarity –
scalers and MCAs generally
require positive polarity
shaped pulses.
Check output pulse polarity
with oscilloscope
Change CP15 internal polarity
jumper setting.
No power to CP15.
Check power LED, check
voltages
Correct power supply.
Polarity jumper not in place
Check that jumper is fitted to
INV or NONINV position
Fit jumper as required.
Missing HV bias voltage to
detector
Check voltage.
Supply correct bias voltage for
the radiation detector.
CP15 input damaged.
Use test input to inject a
signal.
Arrange for repair.
Output pulses distorted
Output saturating.
Check output pulse shape with Use appropriate gain setting.