3

 

Quick Start 

Step 1:

 

Download  the  installer  package  and  the  manual  of  the 

logger software 

Pyro Workbench

 from the 

PyroScience

 website:  

https://www.pyroscience.com/downloads.html

 Unzip and start the 

installer, and follow the instructions.  

Step 2:

 Connect the 

FireSting pro

 meter

 

with the micro-USB cable 

to the Windows PC/laptop (Windows 7, 8, 10). 

Step 3:

 Connect appropriate 

PyroScience

 sensor(s) to the  optical 

sensor connectors 1 to 4 at the device (see 4)

Step 4:

 Connect an external temperature sensor (item no. 

TDIP15

 

or 

TSUB21

)  to  the  Pt100  connector  or,  alternatively,  an  optical 

temperature sensor (see 5) to one of the optical sensor connectors 
1 to 4 for automatic temperature compensation. 

Step 5:

 Prepare appropriate calibration standards, as described in 

the respective sensor manuals (see 7)

Step 6:

 Start the logger software

 

by clicking on the short-cut "Pyro 

Workbench" on your desktop. 

Step 7:

 Open the Settings Wizard by clicking on the 

FireSting pro

 

picture. Select the respective analyte and enter all sensor settings 
for  each  sensor,  including  an  appropriate  mode  of  temperature 
compensation.  

Step 8: 

Open  the  Calibration  Wizard  and  follow  the  calibration 

instructions  for  each  sensor.  Measurements  with  the  respective 
sensors will start automatically after all required sensor calibrations 
have been performed. 

Step 9:

 Configure the Graphs according to your preferences. 

Step 10:

 Activate Data Logging

Summary of Contents for FireSting pro

Page 1: ...1 FireSting pro FIBER OPTIC MULTI ANALYTE METER USER MANUAL ...

Page 2: ...he FireSting pro is manufactured by PyroScience GmbH Hubertusstr 35 52064 Aachen Germany Phone 49 0 241 5183 2210 Fax 49 0 241 5183 2299 Email info pyroscience com Internet www pyroscience com Registered Aachen HRB 17329 Germany ...

Page 3: ...ensor lifetime pH oxygen and temperature determination simultaneously in one sample This single device solution for multi analyte measurements has integrated atmospheric pressure and humidity sensors for precise and easy oxygen sensor calibration but also for automatic pressure compensation of the oxygen measurements Furthermore the FireSting pro offers 4 analog outputs and a broadcast mode The ne...

Page 4: ...or temperature which is currently measured and can be changed for each channel The air inlet equilibrates the internal temperature pressure and humidity sensors with the surrounding Avoid covering these holes to ensure free air circulation towards the internal sensors The micro USB connector on the left side panel provides energy supply and data exchange with the PC Right hand side of it a connect...

Page 5: ... 5 to one of the optical sensor connectors 1 to 4 for automatic temperature compensation Step 5 Prepare appropriate calibration standards as described in the respective sensor manuals see 7 Step 6 Start the logger software by clicking on the short cut Pyro Workbench on your desktop Step 7 Open the Settings Wizard by clicking on the FireSting pro picture Select the respective analyte and enter all ...

Page 6: ... a male fiber plug Remove the black caps from the plug of the sensor fiber Remove the red caps from the sensor ports at the FireSting pro the red caps should be put on again if it is not in use anymore to protect the optics Insert the male fiber plug of the sensor cable into the ST port female fiber connector of the FireSting pro and turn the bayonet coupling gently clockwise until the plug is loc...

Page 7: ...Sting pro is compatible with a broad range of optical oxygen pH and temperature sensors from PyroScience For an overview of available optical sensor types please see the PyroScience website https www pyroscience com products html ...

Page 8: ...rt Range Resolution Accuracy 1 channel for 4 wire Pt100 30 C to 150 C 0 02 C 0 5 C Internal temperature sensor Range Resolution Accuracy 40 to 125 C 0 01 C 0 3 C Internal pressure sensor Range Resolution Accuracy 300 to 1100 mbar 0 1 mbar typ 3 mbar Internal humidity sensor Range Resolution Accuracy 0 to 100 rel humidity RH 0 04 RH typ 0 2 RH Digital interface at extension port X1 7 pins UART with...

Page 9: ...tailed instructions for using the Pyro Workbench and application of optical oxygen pH and temperature sensors Manual for logger software Pyro Workbench Windows manuals for optical sensors from PyroScience oxygen pH temperature ...

Page 10: ...ecially after removal of the protective cap Prevent mechanical stress to the fragile sensing tip Prevent injuries with needle type sensors Calibration and application of the sensors is on the user s authority as well as data acquisition treatment publication The optical sensors and meter FireSting pro are not intended for medical diagnostic therapeutic or military purposes or any other safety crit...

Reviews: