5
3
Quick Start
Step 1:
Download the installer package and the manual of the
logger software
Pyro Workbench
from the
PyroScience
website:
https://www.pyroscience.com/downloads.html
installer, and follow the instructions.
Step 2:
Connect the
FireSting pro
meter
with the micro-USB cable
to the Windows PC/laptop (Windows 7, 8, 10).
Step 3:
Connect appropriate
PyroScience
sensor(s) to the optical
sensor connectors 1 to 4 at the device (see 4).
Step 4:
Connect an external temperature sensor (item no.
TDIP15
or
TSUB21
) to the Pt100 connector or, alternatively, an optical
temperature sensor (see 5) to one of the optical sensor connectors
1 to 4 for automatic temperature compensation.
Step 5:
Prepare appropriate calibration standards, as described in
the respective sensor manuals (see 7).
Step 6:
Start the logger software
by clicking on the short-cut "Pyro
Workbench" on your desktop.
Step 7:
Open the Settings Wizard by clicking on the
FireSting pro
picture. Select the respective analyte and enter all sensor settings
for each sensor, including an appropriate mode of temperature
compensation.
Step 8:
Open the Calibration Wizard and follow the calibration
instructions for each sensor. Measurements with the respective
sensors will start automatically after all required sensor calibrations
have been performed.
Step 9:
Configure the Graphs according to your preferences.
Step 10:
Activate Data Logging.