background image

6

 

Specifications of the FireSting pro 

Dimensions 

68 x 120 x 22 mm (housing) 
78 x 120 x 24 (total) 

Weight 

ca. 290 g 

Interface 

USB 2.0 

Power supply 

USB-powered  (max 50mA at 5V) 

Supported operating systems 

Windows  7, 8, 10 

Operating temperature 

0 to 50°C 

Max. relative humidity 

non-condensing conditions 

Optical sensor port 

1, 2, or 4 

(dependent on model)

 

Optical sensors  
 

complete 

PyroScience

 sensor 

portfolio for O2, pH and T 

Optical sensor connector 

fiber-optic ST-plug 

Max. sample rate 

ca. 10 samples per second 

(depending on the settings)

 

External temperature port 
Range, Resolution, Accuracy * 

1 channel for 4-wire Pt100 
-30°C to 150°C, 0.02°C, 

0.5°C 

Internal temperature sensor 
Range, Resolution, Accuracy * 

-40 to 125°C, 0.01°C, 

0.3°C 

Internal pressure sensor 
Range, Resolution, Accuracy 

300 to 1100 mbar, 0.1 mbar, typ. 

mbar 

Internal humidity sensor 
Range, Resolution, Accuracy 

0 to 100%  rel. humidity (RH), 0.04% 
RH, typ. 

0.2% RH 

Digital interface 
at extension port X1 (7 pins) 

UART with 3.3V levels (5V tolerant), 
115 200 baud, 8 data bit, 1 stop bit, 
no parity, no handshake 

Connector plug port X1 

Phoenix Contact, item no. 1778887 

Analog output (4 channels) 
at extension port X2 (5 pins) 

0 to 2.5 VDC 
14 bit resolution 

Connector plug port X2 

Phoenix Contact, item no. 1778861 

 
* Please note, that the optical sensors have a different temperature range. 

Summary of Contents for FireSting pro

Page 1: ...1 FireSting pro FIBER OPTIC MULTI ANALYTE METER USER MANUAL ...

Page 2: ...he FireSting pro is manufactured by PyroScience GmbH Hubertusstr 35 52064 Aachen Germany Phone 49 0 241 5183 2210 Fax 49 0 241 5183 2299 Email info pyroscience com Internet www pyroscience com Registered Aachen HRB 17329 Germany ...

Page 3: ...ensor lifetime pH oxygen and temperature determination simultaneously in one sample This single device solution for multi analyte measurements has integrated atmospheric pressure and humidity sensors for precise and easy oxygen sensor calibration but also for automatic pressure compensation of the oxygen measurements Furthermore the FireSting pro offers 4 analog outputs and a broadcast mode The ne...

Page 4: ...or temperature which is currently measured and can be changed for each channel The air inlet equilibrates the internal temperature pressure and humidity sensors with the surrounding Avoid covering these holes to ensure free air circulation towards the internal sensors The micro USB connector on the left side panel provides energy supply and data exchange with the PC Right hand side of it a connect...

Page 5: ... 5 to one of the optical sensor connectors 1 to 4 for automatic temperature compensation Step 5 Prepare appropriate calibration standards as described in the respective sensor manuals see 7 Step 6 Start the logger software by clicking on the short cut Pyro Workbench on your desktop Step 7 Open the Settings Wizard by clicking on the FireSting pro picture Select the respective analyte and enter all ...

Page 6: ... a male fiber plug Remove the black caps from the plug of the sensor fiber Remove the red caps from the sensor ports at the FireSting pro the red caps should be put on again if it is not in use anymore to protect the optics Insert the male fiber plug of the sensor cable into the ST port female fiber connector of the FireSting pro and turn the bayonet coupling gently clockwise until the plug is loc...

Page 7: ...Sting pro is compatible with a broad range of optical oxygen pH and temperature sensors from PyroScience For an overview of available optical sensor types please see the PyroScience website https www pyroscience com products html ...

Page 8: ...rt Range Resolution Accuracy 1 channel for 4 wire Pt100 30 C to 150 C 0 02 C 0 5 C Internal temperature sensor Range Resolution Accuracy 40 to 125 C 0 01 C 0 3 C Internal pressure sensor Range Resolution Accuracy 300 to 1100 mbar 0 1 mbar typ 3 mbar Internal humidity sensor Range Resolution Accuracy 0 to 100 rel humidity RH 0 04 RH typ 0 2 RH Digital interface at extension port X1 7 pins UART with...

Page 9: ...tailed instructions for using the Pyro Workbench and application of optical oxygen pH and temperature sensors Manual for logger software Pyro Workbench Windows manuals for optical sensors from PyroScience oxygen pH temperature ...

Page 10: ...ecially after removal of the protective cap Prevent mechanical stress to the fragile sensing tip Prevent injuries with needle type sensors Calibration and application of the sensors is on the user s authority as well as data acquisition treatment publication The optical sensors and meter FireSting pro are not intended for medical diagnostic therapeutic or military purposes or any other safety crit...

Reviews: