4. Operating Environment
The Tester Must Be Located in a Suitable Environment
●
All Q-Lab test chambers are sophisticated scientific instruments.
●
All tester models must be operated in a suitable controlled environment (
).
●
Operating the tester in an unsuitable environment (
) will void the warranty.
4.1 Suitable Environments (May 2020)
Ambient Laboratory Temperature and Humidity
●
The recommended ambient operating temperature and relative humidity (RH) for Q-SUN testers is 23 ± 5 °C and
50 ± 25% RH.
●
Operating outside the recommended range (or in rare cases, even within it), certain standards or test cycle conditions
may not be achievable.
●
Operating outside the recommended range can result in the tester producing chamber temperature and/or humidity
faults.
●
Never operate your tester in lab temperatures >40 °C or >80% RH.
●
Consult with Q-Lab for more specific information about achievable chamber temperature/humidity values based upon
various ambient lab conditions.
Physical Environment
●
A room that is dry, clean, and free of dust, particles, gases, or salt fog.
●
A room with an HVAC (heating/ventilation/air-conditioning) system.
●
A location away from windows or HVAC vents.
●
A location that provides the necessary minimum clearances as specified in
Section 4. Operating Environment