1. Specifications, Classifications, Symbols (Jun 2022)
Specifications, Classifications
●
The recommended ambient operating temperature and relative humidity (RH) for the Q-SUN
®
Xe-2 tester is 23 ± 5
°C
and 50 ± 25% RH.
●
The maximum ambient operating temperature and humidity is 40
°C and 80% relative humidity.
●
Temperatures outside the recommended range may cause chamber temperature and/or humidity faults.
●
Transportation and Storage Temperature: -40 °C to 80 °C.
●
Installation Category: Category II for transient over-voltages.
●
Pollution Control: Pollution Degree 2.
●
Sound Pressure Level: Sound Pressure Level does not exceed 75 dBA.
●
Altitude: 2000 meters or less.
●
Operation: Continuous Rating.
●
Supply Connection: Permanently connected or plug/socket connection (industrial type per IEC 6309 or twist lock type
in North America).
●
External Disconnect: Required for all connections.
●
External Over-Current Protection: Must be rated for not more than 40 A (USA, Canada) or 64 A (Europe).
●
Short Circuit Current Rating (SCCR): 1000 A.