Operating Procedures
5-22
PyroMark Q24 MDx User Manual 01/2016
Note
: Be sure to observe all national, state, and local
environmental regulations for the disposal of laboratory
waste.
Cleaning and testing the reagent cartridge
If the reagent cartridge is to be reused, clean it immediately
after use and ensure that it can be used for analysis. It is
recommended that the reagent cartridge is used a maximum
of 30 times.
WARNING
Sharp needles
Do not touch the sharp needles at the bottom of the
reagent cartridge. Handle the needles with care. Small
particles and fibers may obstruct the needles.
Note
: Be sure to observe all national, state, and local
environmental regulations for the disposal of laboratory
waste.
The following items are required:
Powder-free gloves
High-purity water (Milli-Q
18.2 MΩ x cm or equivalent)
Beaker (not always required)
Lint-free tissues.
To clean and check that the reagent cartridge can be used
for analysis:
1.
Discard any solutions remaining in cartridge.
2.
Rinse the cartridge compartments 4 times with high-
purity water.
3.
Spray the outside of the needles using high-purity water.
4.
Check that the cartridge needles are not blocked or
damaged. Completely fill the compartments with high-
purity water. Hold the cartridge over a sink or beaker
while pressing firmly on top of each compartment with a
finger (wear powder-free gloves). A jet of water should
come straight out of the tip of each needle.
Summary of Contents for PyroMark Q24 MDx
Page 68: ...Maintenance 6 10 PyroMark Q24 MDx User Manual 01 2016 This page intentionally left blank ...
Page 76: ...Troubleshooting 7 8 PyroMark Q24 MDx User Manual 01 2016 This page left intentionally blank ...
Page 82: ...Glossary 8 6 PyroMark Q24 MDx User Manual 01 2016 This page intentionally left blank ...
Page 98: ...Appendix B B 16 PyroMark Q24 MDx User Manual 01 2016 This page intentionally left blank ...
Page 100: ...Appendix C C 2 PyroMark Q24 MDx User Manual 01 2016 This page intentionally left blank ...
Page 103: ......