QIAcuity User Manual 06/2022
285
Appendix A – Technical Data
QIAGEN reserves the right to change specifications at any time.
Operating conditions
Power
100–240 V AC, 50/60 Hz, Mains supply voltage fluctuations are not to
exceed 10% of nominal supply voltages.
Maximum power consumption:
QIAcuity One, 2plex: 1000VA
QIAcuity One, 5plex: 1000VA
QIAcuity Four: 1000VA
QIAcuity Eight: 1500VA
Fuse
2x T12A L 250V
Overvoltage
category
II
Air temperature
15–32ºC (59.0–89.6ºF)
Relative humidity 10–75% (non-condensing)
Altitude
Up to 2000 m (6500 ft.)
Place of
operation
For indoor use only
Pollution level
2
Environmental
class
3K21 (IEC 60721-3-3)