QIAsymphony SP/AS Consolidated Operating Guide 12/2017
170
13.5.3
Tip waste errors
Error
Comments and suggestions
Tips are stacking in the tip
chute.
Ensure that the tip disposal bag is empty and that it is not jammed between the drawer and
the workbench.
Tips are spilled in the lab.
Ensure that the tip disposal bag is correctly attached to the waste bag holder.
13.5.4
Maintenance errors
Error
Comments and suggestions
Interruption of a
maintenance protocol.
It is not possible to stop a maintenance protocol after it has been started.
Hood open.
During some maintenance protocols the hood might be open. Ensure that the hood is closed
afterwards.
Wrong cleaning agents. Only use recommended cleaning agents. Use of cleaning agents that are not recommended
may result in damage to the QIAsymphony SP/AS instruments.
13.5.5
Configuration
menu errors
Error
Comments and suggestions
The adapter for AS is not
displayed in the
configuration dialog.
Ensure that you have transferred the adapter file(s) to the
Labware AS
folder.
13.6
QIAsymphony SP errors that do not have error codes
13.6.1
“Eluate” drawer
Error
Comments and suggestions
Filter-tips are bent or
deformed after eluate
transfer.
Be sure to define the correct type of eluate rack on the corresponding elution slot. Make sure
that the elution rack is correctly positioned on the elution slot. Only use elution racks that are
compatible with the specified adapter.
The inventory scan of the
“Eluate” drawer detects
an elution rack on
“Elution slot X” although
no elution rack has been
placed there.
If there is an adapter on the elution slot, be sure to remove it. Do not expose the instrument to
direct sunlight (see Section 4.2 of the
QIAsymphony SP/AS User Manual — General
Description
).
Ensure that the bar codes on the drawer can be easily read.
If the error persists, contact QIAGEN Technical Services.
The bar code of the
elution slot/elution rack
cannot be read using the
handheld bar code
scanner.
Make sure that the handheld bar code scanner is properly connected to the QIAsymphony SP.
Try to read other bar codes with the scanner. Make sure that all bar codes can be easily read.
Define the elution slot/elution rack by manually entering the information into the touchscreen.
Tips/channels are
incorrectly positioned on
the elution slot during the
elution step.
Make sure to place the elution rack onto the elution slot in the correct orientation. Be sure to insert and
to define the same sample tube. Only use compatible sample tubes/ racks. For more information about
tubes and racks, visit www.qiagen.com/goto/QIAsymphony.
If the error persists, contact QIAGEN Technical Services.
Summary of Contents for QIAsymphony SP
Page 213: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 213 This page intentionally left blank...
Page 214: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 214 This page intentionally left blank...
Page 216: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 216 www qiagen com...