QIAsymphony SP/AS Consolidated Operating Guide 12/2017
62
If the protocol requires eluate cooling or if you are planning to set up an integrated run, make
sure to use slot 1. Place the elution rack into the appropriate cooling adapter.
Important
Elution rack cooling can be turned off by pressing the snowflake button to the left
of “Elution slot 1”. We do not recommend turning off elution rack cooling if this
is required by the protocol.
7.
An adapter may be required depending on the elution rack being used.
Select the elution rack type from the list. Use the up and down arrows to scroll through the list.
Important
The QIAsymphony SP provides automated assignment of elution racks. If you are
using an Elution Microtube Rack (EMTR), scan the bar code of the rack and the
elution rack type will be automatically selected by the QIAsymphony SP.
Important
If the elution rack was used in a previous run, cooling will be switched on by the
QIAsymphony SP automatically when the next batch requiring eluate cooling is
ordered.
Important
The maximum number of columns that can be reserved depends on the rack size
and on the batches that are already queued for that slot.
Important
: If there are already eluates in an elution rack from a previous run loaded in the
“Eluate” drawer, ensure that the cooling temperatures are suitable for these eluates on the same
elution rack. Otherwise, the eluates may be given the status “invalid”. The QIAsymphony SP
cannot detect whether cooling temperatures are suitable for eluates that are already loaded from
previous runs in the “Eluate” drawer.