QIAsymphony SP/AS Consolidated Operating Guide 12/2017
75
Example of insert for tube carrier.
Important
Place the tubes into the tube carrier in a way that all bar codes are oriented to
the left so that they can be read by the bar code reader.
Important
Depending on the
Configuration Profile
, only bar code labeled sample tubes can
be used. If other tubes are used, a batch or run cannot be defined.
The instrument detects tube size by reading the bar code on the insert or on the tube carrier. If a
tube is used that is not the default tube type for a certain insert, the user must specify the tube type
when defining the sample batch. Default tubes can also be configured.
Inserting samples using a tube carrier
1.
Open the “Sample” drawer by pulling the door toward you.
Five slots are available: The first 4 slots can accommodate tube carriers containing sample
tubes; the fifth slot “A” accommodates a tube carrier containing internal control.
The status of each slot is shown by LEDs located behind the stop line. The LEDs may be
illuminated in green, orange, or red.
Green — slot is free and ready for loading
Orange — tube carrier is loaded
Red — slot is currently locked
Summary of Contents for QIAsymphony SP
Page 213: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 213 This page intentionally left blank...
Page 214: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 214 This page intentionally left blank...
Page 216: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 216 www qiagen com...