QIAsymphony SP/AS Consolidated Operating Guide 12/2017
80
If the tubes are not bar code labeled, information about the internal control must be entered
manually.
After insertion of the tube carrier into “Slot A”, follow the steps below to enter information about
the internal control.
1.
Press the
IC
button to check or modify the internal controls.
The
Internal Controls
screen appears.
2.
Select the position that needs an internal control to be manually
assigned by pressing the button.
3.
Select the internal control in the
Internal controls:
list.
4.
Press
OK
to assign internal controls to the selected positions.
5.
Press
OK
to confirm the overall assignments of internal control.