Figure #8
4.6.1.4
Set the dissolution
tester to 25 RPMs. Start the spindle rotation and
observe the dial indicator travel. Rotate the dial face
so that the minimum counter clockwise pointer
position coincides with “0” on the indicator. As the
shaft rotates, the pointer will move clockwise to a
maximum value. This distance is the total indicated
reading (T.I.R.) or wobble. Each division on the dial
indicator is equal to .01mm.
4.6.2
For VanKel / Varian
Testers
4.6.2.1
Remove the centering
ring, dissolution vessel and assemble the dial
indicator to the bracket assembly as shown in figure
#5.
4.6.2.2
Place the indicator
assembly onto the centering ring guide post as shown
in figure #6.
4.6.2.3
Repeat sections 4.6.1.3
& 4.6.1.4
4.7
Measuring
Vessel Centering
4.7.1
For Testers with One-
Piece Shafts