Page
16
of
18
MANVALTOL-KT2 Rev. A, 6-15-2010
the dial face so that the minimum counter clockwise
pointer position coincides with 40 on the indicator.
4.7.1.7
As the shaft rotates, the
pointer will move clockwise to a maximum value. This
distance is the total indicated reading or T.I.R.. Each
division on the dial indicator is equal to .05mm.
4.7.1.8
Stop the spindle
rotation and raise the drive head so that the indicator
ball is near the top of the vessel just below the rim
and repeat sections 4.7.1.5 to 4.7.1.7
4.7.2
For Testers with Two-
Piece, Spin-Off Shafts
4.7.2.1
Follow the instructions in section 4.7.1 with the
following exceptions. Remove only the lower portion
of the paddle or basket shaft assembly and replace it
with one of the following Spin-Shaft adapters as
shown in figure #11. (Distek Testers do not require
the use of an adapter.)
Figure #11
Spin Shaft Adapter
for VanKel / Varian
Spin Shaft Adapter
for Hanson / Distek