37 MAINTENANCE
3. Cleaning and Disinfection
BIOLOG 3000
i
can be cleaned and disinfected very simply, as follows:
To clean the BIOLOG 3000
i
plastic case, wipe the surfaces of the case
with a clean cloth moistened with water only.
To clean the display, gently wipe the plastic display lens with a soft,
damp cloth moistened with water or a commercial glass cleaner.
NOTE:
Do not wipe the case or the display lens with alcohol. Alcohol will damage plastic
surfaces over time, and may cause them to crack. In addition, the display lens may become
'foggy' in appearance.
To disinfect the BIOLOG 3000
i
, wipe the back of the device, including
the electrodes, with a hospital grade disinfectant.
NOTE:
Do not saturate the BIOLOG 3000
i
with any liquid because excess fluid could enter the
case and damage the electronics.