background image

Patient Preparation

•  Select area over bone, avoiding muscle areas and breast tissue.
•   Clean electrode sites aggressively with alcohol and use a razor 

to remove hair.

•  Abrade electrode sites with a scrub pad or gauze. Allow sites  

to dry before applying electrodes.

•  Firmly snap electrodes to the lead wires.
•  Apply electrodes to appropriate sites. Press 

adhesive border firmly for consistent adhesion.

•  To help alleviate tension, loop and tape lead 

wires down. 

Electrode

Tape

5-Electrode Placement

 

   Electrodes

      –        +

                       

       RED      BROWN

       RED      BLACK

     WHITE    BLACK

           

    GREEN (Ground)

Ch. 1

CMS

Ch. 2 

aVF

Ch. 3

lead II

Q200/HE Hook-Up Guide 
for Holter

QRS Diagnostic 

 

6901 E. Fish Lake Road, Suite 188 
Maple Grove, MN, USA 55369
800.465.8408    |    +1.763.559.8492    |    Fax +1.763.559.2961
info@QRSdiagnostic.com    |    www.QRSdiagnostic.com

0086

Release Date: 04/12

Advena Ltd

33 Bridge St.

Hereford HR4 9DQ

UK

© QRS 

600005-00A   

Reviews: