background image

 

 

 

Models SSC-2 & SSC-4 

SOURCE 

STORAGE CONTAINERS 

 

 

 

User Instructions

MAN-056-December 2015 

 

Summary of Contents for Sentinel SSC-2

Page 1: ...Models SSC 2 SSC 4 SOURCE STORAGE CONTAINERS User Instructions MAN 056 December 2015...

Page 2: ...ther than the source connector during hook up to the drive cable the radiation emitting source assembly must never be touched by hand or handled unshielded up close Special confirmatory surveys must b...

Page 3: ...e resulted from the use of such product in accordance with instructions given by or referred to by the manufacturer QSA Global Inc assumes no liability or responsibility for the usage of any radioacti...

Page 4: ...nges users within the USA must be specifically licensed to do so Applications for a license should be filed with the Materials Licensing Section of the appropriate U S Nuclear Regulatory Commission re...

Page 5: ...2 USER INSTRUCTIONS 3 2 1 Opening unlocking the storage container 3 2 2 Loading sources into the storage container 3 2 3 Unloading source from the storage container 4 2 4 Closing locking the storage...

Page 6: ...um DU Lead with DU insert Weight 120 pounds 54 kg 165 pounds 75 kg Capacity Isotope 2 x 120 Curies Ir 192 Se 75 4 x 100 Curies Ir 192 Se 75 Overall Diameter 15 inches 38 cm Overall Height 19 inches 48...

Page 7: ...TECHNICAL SPECIFICATION Figure 1 Cutaway view of the Storage Container SSC 4 Shield Pot Shown 2...

Page 8: ...container 2 1 Opening unlocking the storage container 1 Perform a confirmatory radiation survey on and around the storage container to ensure radiation measurements are acceptable 2 If radiation measu...

Page 9: ...Open the storage container according to Section 2 1 2 Locate the source to be unloaded from the inner shielded pot 3 Follow the directions in the exposure device operating manual for source transfers...

Page 10: ...tep 3 Inspect the contents for damage or missing components Refer to Figure 1 4 Attach the cover and cover clamp band 5 Install the lock bolt into the cover clamp band Ensure the clamp band is tight e...

Page 11: ...r to its first use after removal from storage Acceptable results of a radio assay must indicate removable contamination in less than 185 Bq 0 005 Ci Evaluation of depleted sources for return to QSA Gl...

Page 12: ...ipment Damaged cropped modified or contaminated source assemblies may require special handling and special transport containers Notify the authorized recipient for specific instructions in these circu...

Page 13: ...3 2216 Sales Service SENTINEL Loma Systems QSA Global Inc U Lomy 1069 334 41 Dobrany Czech Republic Telephone 420 377 183 843 Email SENTINELSales QSA GLOBAL com Website www qsa global com All goods an...

Reviews: