Appendix A – Technical Specifications
146
www.qualitrolcorp.com
Standards
References/Test criteria
inputs( V ), Relay
Passed with : Criteria A
IEC 61000-4-12 Ring wave test
Tested to IEC 61000-4-12, Tested for Level
3,+/- 2 kV Common Mode, 1 kV Differential
Mode, 100KHz
Tested on Power-line, Digital, Analog
inputs( V ), Relay
Passed with : Criteria A
IEC 61000-4-5 Surge test
Tested to IEC 61000-4-5 ,Tested for Class
4, I/O 4 KV Common Mode, 2KV Differential
Mode
Tested on Power-line, Digital, Analog
inputs( V ), Relay
Passed with : Criteria A
ENV 50204 Immunity to EMI from
digital radio telephones
Tested to ENV 50204, Tested for level 3, 10
V/m, @900 MHz +/- 5 MHz, @1890 MHz +/-
10 MHz keyed at frequency 200 Hz +/- 1%,
50% duty cycle [2.5 ms ON 2.5 ms OFF].
Tested on Complete Device
Passed with : Criteria A
IEC 60255-5 Clause 6 Dielectric test
ANSI IEEE C37.90 – 1989 SWC
dielectric tests
Tested to 2.5kV for 1 minute, Tested for 2.5
kV for Power-port, Analog inputs (V &
I),Digital inputs, Relay, and 1 kV for isolated
communication ports (RS232 & RS485) .
No flashover or breakdown occurred.
IEC 60255-5 Clause 7 Insulation test
Tested to applied 500 VDC with respect to
Earth.
Tested on Power port, Analog inputs (V & I),
Digital inputs & Relay.
The impedance was greater than
100Mohms.
IEC 60255-5 Clause 8 Impulse voltage
test
Tested to +/-5 kV, 0.5 J.
Tested on: Power-line, digital, Analog inputs
(V & I) & Relay.
No flashover or breakdown occurred.
IEC 61000-4-11 Supply Voltage Dips
Tested to Dips-Class-3,Interruptions-Class-3
Summary of Contents for INFORMA PMD-A
Page 1: ...User Manual www qualitrolcorp com INFORMA PMD A Data Acquisition Unit Document ID 40 08557 01...
Page 7: ...Contents Contents iv www qualitrolcorp com References 240...
Page 8: ...About This Manual 1 www qualitrolcorp com Chapter 1 About This Manual...
Page 12: ...Overview 5 www qualitrolcorp com Chapter 2 Overview...
Page 15: ...Components and Specifications 8 www qualitrolcorp com Chapter 3 Components and Specifications...
Page 58: ...Installation 51 www qualitrolcorp com Chapter 4 Installation...
Page 86: ...Configuration 79 www qualitrolcorp com Chapter 6 Configuration...
Page 108: ...Configuration 101 www qualitrolcorp com Figure 6 22 PQ Free Interval Continuous Recording...
Page 120: ...Maintenance 113 www qualitrolcorp com Chapter 7 Maintenance...
Page 126: ...Maintenance 119 www qualitrolcorp com Figure 7 7 Voltage Calibration...
Page 166: ...Appendix B INFORMA PMD A Functions 159 www qualitrolcorp com Figure B 9 RMS Envelope PQRVC...
Page 199: ...Appendix D License Information 192 www qualitrolcorp com Appendix D License Information...
Page 246: ...References 239 www qualitrolcorp com References...