Event Code
Message Text and Description
Details and Solution
4090
Auto calibration of one or more modules
failed. Library not properly calibrated. This
may cause media movement failures.
The library needs to be re-calibrated. Ensure the
library firmware is up to date. This event indicates
that one or more of the gray calibration targets located
on the library magazines could not be used in
calibration. Inspect the calibration targets in each
module and then initiate the auto-calibration routine
via
Maintenance > Auto Calibration
.
4091
Auto calibration of one or more modules
failed. Library not properly calibrated. This
may cause media movement failures.
The library needs to be re-calibrated. Ensure the
library firmware is up to date. This event indicates
that one or more of the gray calibration targets located
on the library magazines could not be used in
calibration. Inspect the calibration targets in each
module and then initiate the auto-calibration routine
via
Maintenance > Auto Calibration
.
4092
Installed robotic does not support auto
calibration.
If move errors are occurring, the robotic assembly
needs to be replaced with a robot that supports auto-
calibration. If this event persists after replacing the
robot, manually initiate calibration via
Maintenance >
Auto Calibration.
4093
Could not obtain an IP address from DHCP
server.
Check the network configuration settings and check if
the DHCP server is reachable. Use the network
configuration menu or unplug the network cable and
plug it in after a few seconds to trigger an automatic
reconfiguration of the network interface.
4094
Drive interface I/O error.
Reboot the library for initialization of hardware and
device drivers. If the problem persists, contact
Technical Support.
4095
Library test failed. Not enough valid
cartridges available for testing.
Add at least 1 valid tape cartridge to the library.
4098
System time synchronization via SNTP
failed.
Check for valid SNTP server address in Time
configuration. If correct, ensure that server reachable
from your network and not blocked by Firewall.
4099
An unexpected reset of robotics has been
detected.
Ensure that the spooling cable is fully seated in the
Base Module and connected correctly to the robotic
assembly. If the error re-occurs replace the robotic
assembly.
4110
Drive disabled due to an incompatible Drive
Power Board.
Remove incompatible Drive Power Board. Only install
Drive Power Boards that are compatible with the
library.
4113
Move Cartridge operation failed due to
cartridge not properly taken over from
Drive.
Check for labels or cartridge misalignments that would
prevent the cartridge from coming out of the slot or
drive.