6-4
Troubleshooting
501510 Rev. S
that there is not a loose cartridge in the Handler (if a cartridge is in the Handler, and
it is properly gripped, this will not cause a problem.) A failed infrared emitter or de-
tector could also be responsible. A large piece of dust or an errant barcode label could
also cause this failure.
6.2.9
Z Failed to Home
During POST (Power On Self Test) the RLS did not detect that the Handler’s Z
-Axis
(insertion) successfully found its home position. Check that there are no loose maga-
zines or cartridges in the library. Check that the drives are properly seated and the
locking lever is fastened. Check that there is not a loose cartridge in the Handler (if a
cartridge is in the Handler, and it is properly gripped, this will not cause a problem.)
A failed infrared emitter or detector could also be responsible. A large piece of dust or
an errant barcode label could also cause this failure.
6.2.10
No Drivebay Found
During POST (Power On Self Test) the Executive microprocessor could not communi-
cate with Drive Bay microprocessor. Check that the 10-conductor ribbon cable be-
tween the Interconnect and Drive Bay PCBAs is properly installed. Go to the
Pri-
vate\Cabling
menu to see if any cables are unplugged.
6.2.11
No Carriage Found
During POST (Power On Self Test) the Executive microprocessor could not detect the
presence of the Carriage Assembly. Check the 50-conductor cables between the Exec-
utive, Interconnect, and 2-Axis Carriage PCBAs. Go to the
Private\Cabling
menu to see if any cables are unplugged.
6.2.12
No Sensor Pcb Found
During POST (Power On Self Test) the Executive microprocessor could not detect the
presence of the Sensor Master PCBA. Check the cable from the Interconnect PCBA to
the Sensor Master PCBA. Go to the
Private\Cabling
menu to see if any cables
are unplugged.
6.2.13
No Carousel Found
During POST (Power On Self Test the Executive microprocessor could not detect the
presence of the Carousel PCBA or could not communicate with its microprocessor.
Check the cable between the Drive Bay and the Carousel PCBAs. Go to the
Pri-
vate\Cabling
menu to see if any cables are unplugged.
6.2.14
No Storage Array
During POST (Power On Self Test) the Executive microprocessor could not detect the
presence of the Storage Array or could not communicate with its microprocessor.
Summary of Contents for RLS-4124
Page 1: ...RLS 8000 Tape Library Technical Service Manual 501510 Rev S...
Page 2: ......
Page 14: ...xiv 501510 Rev S This page left blank intentionally...
Page 16: ...1 2 Introduction 501510 Rev S Table 1 1 Applicable Documents...
Page 20: ...1 6 Introduction 501510 Rev S This page left blank intentionally...
Page 23: ...501510 Rev S Description and Theory of Operation 2 3...
Page 34: ......
Page 50: ...3 16 The Operator Interface 501510 Rev S This page left blank intentionally...
Page 64: ...4 14 The Maintenance Menu 501510 Rev S This page left blank intentionally...
Page 65: ...501510 Rev S The Private Menu 5 1 5 The Private Menu...
Page 69: ...501510 Rev S The Private Menu 5 5 10 Close the Front Panel Door...
Page 188: ...8 10 Firmware Updating 501510 Rev S Figure 8 7 Properties Screen...
Page 205: ...501510 Rev S RLS Expansions 9 13 This page left blank intentionally...