SRV02 Gyroscope User Manual

1. Measure Encoder Input Channel #1. 
2. Rotate the gyroscope module inner blue frame, one quarter of a full rotation and the encoder 

should measure 1024 counts in quadrature mode. 

Note

: Some data acquisition systems do not measure in quadrature and, in this case, one-quarter 

of the expected counts are received, i.e. 256 counts in the SRV02-E or 512 in the SRV02-EHR. 
In addition, some data acquisition systems measure in quadrature but increment the count by 
0.25 (as opposed to having an integer number of counts). Make sure the details of the data-
acquisition system being used is known. The counters on the Quanser HIL boards measure in 
quadrature and therefore a total of four times the number of encoder lines per rotation, e.g. a 
1024-line encoder results in 4096 integer counts for every full rotation.

6.2.2. Troubleshooting

If the encoder is not measuring properly, go through this procedure:

Check that the data-acquisition board is functional, e.g. the LED on the Quanser Q2-USB 
terminal board should be bright green. If not, then the terminal board or DAQ board fuses may 
be burnt and need replacement.

Check that both the A and B channels from the encoder are properly generated and fed to the 
data-acquisition device. Using an oscilloscope, there should be two square waves, signals A and 
B, with a phase shift of 90 degrees. If this is not observed then the encoder may be damaged and 
need to be replaced. Please see Section 7 for information on contacting Quanser for technical 
support.

7. Technical Support

To obtain support from Quanser, go to  

http://www.quanser.com/

 and click on the 

Tech Support 

link. 

Fill in the form with all the requested software and hardware information as well as a description of the 
problem encountered. Also, make sure your e-mail address and telephone number are included. Submit 
the form and a technical support representative will contact you.

8. References

[1] Quanser. 

SRV02 User Manual

[2] Quanser. 

Amplifier User Manual

[3] Quanser. 

Data Acquisition Board User Manual

Document Number 908 

 Revision 1.1 

 Page 10

Summary of Contents for SRV02 Series

Page 1: ...Rotary Motion Servo Plant SRV02 SRV02 Gyroscope User Manual...

Page 2: ...nc All rights reserved This work may not be translated or copied in whole or in part without the written permission of the copyright holder except under the terms of the associated software license ag...

Page 3: ...2 5 Support springs 4 2 2 6 Gear mechanism 4 2 2 7 Gyroscope module encoder connector 4 2 2 8 SRV02 plant 4 2 2 9 Support base plates 4 2 2 10 Hold clamps 4 2 2 11 Gyroscope rotation disk motor conne...

Page 4: ...SRV02 Gyroscope User Manual 6 2 Encoder 10 6 2 1 Testing 10 6 2 2 Troubleshooting 10 7 TECHNICAL SUPPORT 10 8 REFERENCES 10 Document Number 908 Revision 1 1 Page iii...

Page 5: ...the external frame in order to provide tension for the inner frame structure The external frame has a hole at its base through which it can be mounted on the output shaft of a SRV02 plant This allows...

Page 6: ...mponents are then described in Section 2 2 2 1 Component Nomenclature The main components of the rotary gyroscope system are labeled below in Figure 2 and listed in Table 1 Please use the ID to locate...

Page 7: ...angular momentum for rotational dynamics experiments by spinning about its spin axis 2 2 2 DC motor This component is depicted by 2 in Table 1 and Figure 2 and is used to rotate the disk component 1...

Page 8: ...and rotor disk structure to the shaft to which the encoder is connected to This mechanism introduces a 1 4 gear ratio from the frame shaft to encoder shaft that has to be taken account when converting...

Page 9: ...yro module inertia about input axis Jg 0 002 Kg m Spring stiffness Ks 1 9089e3 N m Gear mechanism ratio Kg m m Encoder resolution in quadrature mode Kenc 4096 counts rev Table 2 SRV02 and gyroscope sy...

Page 10: ...le Nomenclature Table 3 provides a listing of the standard cables shipped with and used in the wiring of the SRV02 and gyroscope system Cable Designation Description RCA to RCA This cable connects an...

Page 11: ...rd and the amplifier The connections are illustrated in Figure 3 and summarized in Table 4 The details of the wiring procedure are given thereafter Caution Make sure the amplifier and PC are turned OF...

Page 12: ...se consult the appropriate documentation to ensure that the board is installed properly 2 Connect the 5 pin DIN to RCA cable between the terminal board analog output 0 connector and the amplifier From...

Page 13: ...correctly by going through this procedure 1 Apply a voltage to analog output channel 0 of the terminal board 2 The motor gear should rotate counter clockwise when a positive voltage is applied and cl...

Page 14: ...t the data acquisition board is functional e g the LED on the Quanser Q2 USB terminal board should be bright green If not then the terminal board or DAQ board fuses may be burnt and need replacement C...

Reviews: