10
5.3.2.13 Press
Program
and
Down-arrow
at the same time:
This enters the
setpoint-programming mode
.
Display toggles between [
SP_1
] and [
50000
]. 18.xxx for the new model
5.3.2.14 Press the Program button repeatedly until the
Display toggles between [
SP_5
] and [
30000
],10000 for the new model.
Change
SP_5
to
1
.000
Note: If SP_5 is unavailable, perform the next step (SPC_5) then come back
to this step; Ensure SPC_5 is set to [000].
You should now be able to set up SP_5 to 1.000. Continue and perform the
next step before locking the meter.
5.3.2.15 Press Program button repeatedly until the
Display toggles between [
SPC_5
] and [
000
].
Change
SPC_5
to
46
.
5.3.2.16 Check all the setting before lock
STEP B LOCK CODE SETUP:
5.3.2.17 Loosen the meter from cabinet.
5.3.2.18 Remove black face meter bezel.
5.3.2.18 Set both
DIP
Switches from
OFF
to
ON
position.
5.3.2.20 Reinstall the bezel and secure the meter.