Quick Start Guide For Your ULTRA Laser System 

ULTRA Laser Head/Rack Power Supply 

CONGRATULATIONS! 

  

Your system is ready for full operation.   

 

CONSULT CHAPTER 2 OF THE USER’S  

MANUAL BEFORE CONTINUING FOR  

APPROPRIATE EYE PROTECTION AND  

SAFETY PRECAUTIONS. 

 
1. Ensure that the red emergency stop button located on 

the front of the power supply is rotated to the OUT po-
sition. 

2. Insert the key and turn it to the ON position. Note:  

The 

front panel will illuminate and the pump will turn on.

 

3. Check the Pulse Repetition Frequency for the desired 

value and adjust accordingly. 

4. Open the manual shutter on the laser head. 

 
5. Press the RUN/STOP button.  Note:  

You should hear a 

continuous “chirping” as the lamp flashes.

 

6. Press the Q-Switch Button.   
 
Note:  

You can control the output energy with the output 

energy control buttons.

 

 
 
 
 
 
 
 
 
 

 

FOR TECHNICAL SUPPORT  

PLEASE CALL  

CUSTOMER SERVICE: 

QUANTEL USA:  1-800-914-8216 

QUANTEL:  (33)1-6929-1700 

DOC00073 Rev A 

START UP

 

Your laser is pre-configured to operate at full en-
ergy and Pulse Repetition Frequency (PRF).  
 
The only routine maintenance required on the UL-
TRA system is deionization cartridge and flash-
lamp replacement.  Cycle the coolant at least 30 
minutes every 2 weeks and replace the flashlamp 
as required.  See the User's Manual for details. 
 
If you need to ship the laser for any reason, see 
Chapter 3 of the User's Manual for details on 
draining the system for transport to prevent freez-
ing. 

RUN/STOP Button 

Q-Switch Button 

Output Energy Control Buttons 

Optional External DI Cartridge Connection: 
1.Connect the connectors and tubing to the DI car-
tridge. 

2.Connect the connector sides of the DI cartridge to 
the vent/fill connectors located on the back of the 
power supply.  

Shutter CLOSED 

Shutter OPEN 