Bit Error Rate Tester | BERT 1005
Quantifi Photonics Ltd.
Version
2.04
26
6.3.3
Upgrading a MATRIQ instrument with the latest firmware
The Cohesion Operator can upgrade MATRIQ instruments that currently use firmware package
version 2.08.24 or greater. To upgrade MATRIQ instruments that use older software packages,
refer to the user manual that came with your instrument.
To upgrade a MATRIQ instrument:
1.
Get the latest MATRIQ firmware package CohesionMTRQ-<version>.tgz, for example by
downloading it from quantifiphotonics.com (go to Resources > Drivers, software and manuals >
MATRIQ Series), and save it to your network.
2.
Open the Cohesion Operator, for example by double-clicking the Cohesion Operator desktop
icon.
3.
Select the instrument by entering its IP address or by selecting it from the Discovery drop down
list.
4.
To confirm that you have selected the correct instrument, click Connect.
This will retrieve instrument information, with Installed Package showing the current firmware
version.
5.
In Package, click the Browse button, navigate to the previously downloaded firmware package
and select it.
6.
Click Upgrade.
The instrument will be upgraded to the selected firmware package. This can take a few minutes
and the instrument might reboot several times in the process.
7.
A message shows when the upgrade is complete.
To confirm the new firmware version, click Connect to retrieve the latest instrument
information.
If an upgrade attempt is unsuccessful, the Cohesion Operator will stop the upgrade process
and restore the MATRIQ instrument to its previous firmware version. Messages will be
displayed accordingly.