no modules are displayed in the CohesionUI window. This is useful for users who are operating in a
multi-chassis MXI setup, instead of the standard PXIe embedded controller setup.
There are two actions in the SYSTEM controls section:
•
SYNC: Synchronize CohesionUI with the latest information from the CohesionSCPI service.
•
RE-INIT: Re-initialize CohesionUI by synchronizing the CohesionSCPI service with the
CohesionDriver service.
The SYNC button is also displayed on the HOME page beside every chassis in the setup. This allows any
chassis to be synchronized independently. After clicking the SYNC button, CohesionUI will disable the
page while it is synchronizing with the CohesionSCPI service. Once it is complete, the page will be
functional again.
Clicking the RE-INIT button will bring up a prompt to continue, since this action will temporarily
disconnect all modules while the re-initialization with the CohesionDriver service is in progress. Once it is
complete, the page will be functional again.
The RE-INIT action will disconnect any connected users to the PXIe system while the action is
being completed. All modules will be disabled during this time.