Bit Error Rate Tester | BERT 1005
Quantifi Photonics Ltd.
Version
2.04
64
Syntax
:OUTPut<n>:CLOCk:VCOLock?[<wsp>INFO]
Description
Query the stability of the clock synthesizer.
Parameters
INFO: Describes what the return values represent
Response
The status of the clock stability, 1 for a locked state and 0 when there is no lock
established.
Example
:OUTPut5:CLOCk:VCOLock? ->1
:OUTPut5:CLOCk:VCOLock? INFO -> 0: NO LOCK 1: LOCKED
Command
:OUTPut<n>:CLOCk:DIVIde?
Syntax
:OUTPut<n>:CLOCk:DIVIde?[<wsp><MIN|MAX|DEF|SET|LIST|ALL>]
Description
Query the clock synthesizer divide ratio.
Parameters
MIN: Query the minimum configurable value
MAX: Query the maximum configurable value
DEF: Query the default power-on value
SET: Query the current SET value
LIST: Query the LIST of discrete options for configurable values (if available)
ALL: Query the values specified by all parameters
By default, the command Query the current SET value.
Response
A single value or a comma separated string of values. For ALL, the format is
"<MIN>,<MAX>,<DEF>,<SET>,<LIST>".
Example
:OUTPut3:CLOCk:DIVIde? -> 2
Command
:OUTPut<n>:CLOCk:DIVIde
Syntax
:OUTPut<n>:CLOCk:DIVIde<wsp><value|MIN|MAX|DEF>
Description
Set the clock synthesizer divide ratio.
Parameters
value: Set to the desired new set point
MIN: Set to the minimum value
MAX: Set to the maximum value
DEF: Restores to the default value
Response
No response
Example
:OUTPut3:CLOCk:DIVIde 4
Command
:OUTPut<n>:CLOCk:FREQuency:ARBitrary?
Syntax
:OUTPut<n>:CLOCk:FREQuency:ARBitrary?[<wsp><MIN|MAX|DEF|SET|ALL|STEP|
UNIT>]
Description
Query the synthesizer frequency arbitrary settings.
Parameters
MIN: Query the minimum configurable value
MAX: Query the maximum configurable value
DEF: Query the default power-on value
SET: Query the current SET value
ALL: Query all the values specified by the MIN, MAX, DEF, and SET parameters
STEP: Query the minimum resolution of the frequency value
UNIT: Query the default unit for the frequency value
By default, this command Query the current SET value.
Response
A single value or a comma separated string of values. For ALL, the format of the
response is "<MIN>,<MAX>,<DEF>,<SET>".
If the frequency is set to a standard value, the SET value in the response is NAN.
Example
ARBITRARY MODE:
:OUTP3:CLOC:FREQ:ARB? -> 7.96
STANDARD MODE:
:OUTP3:CLOC:FREQ:ARB?-> NAN
Command
:OUTPut<n>:CLOCk:FREQuency:ARBitrary
Syntax
:OUTPut<n>:CLOCk:FREQuency<wsp><value|MIN|MAX|DEF>