Syntax
:SENSe<n>:MEASure<m>:BATHtub:VOLTage<wsp><value|DEF>
Description
Set the bathtub voltage
Parameters
value: The desired new value
DEF: Restores the default value
Response
No response
Example
:SENSe3:MEASure2:BATHtub:VOLTage 7
Command
:SENSe<n>:MEASure<m>:EALarm:MODE?
Syntax
:SENSe<n>:MEASure<m>:EALarm:MODE?[<wsp><DEF|SET|LIST|ALL|INFO>]
Description
Query the mode of ED measurement.
Parameters
DEF: Query the default power-on value
SET: Query the current SET value
LIST: Query the LIST of discrete options for configurable values
ALL: Query the values specified by DEF, SET, and LIST
INFO: Provides a description of what the return values represent
By default, the command Query the current SET value.
Response
A single integer value, or a comma separated integer array. When ALL is requested,
the format is "<DEF>,<SET>,<LIST>".
Example
:SENSe3:MEASure2:EALarm:MODE? -> 0
:SENSe3:MEASure2:EALarm:MODE? INFO ->
0:ACCUMULATED
1:INSTANTANEOUS
Command
:SENSe<n>:MEASure<m>:EALarm:MODE
Syntax
:SENSe<n>:MEASure<m>:EALarm:MODE<wsp><1|INSTantaneous|0|ACCUmulated|
DEF>
Description
Set the mode of the ED measurement.
Parameters
value: The desired new set point
INSTantaneous: Set the mode to one second intervals
ACCUmulated: Set the mode to continually record data until it is stopped
DEF: Restores the default value
Response
No response
Example
:SENSe3:MEASure2:EALarm:MODE INST
Command
:SENSe<n>:MEASure<m>:EALarm:POINt?
Syntax
:SENSe<n>:MEASure<m>:EALarm:POINt?[<wsp><DEF|SET|INFO|LINKED>]
Description
Query the mode of the ED point measurement.
Parameters
DEF: Query the default value
SET: Query the set value
INFO: Provides a description of what the return values represent
LINKED: Query the linked channels
By default, the command Query the current SET value.
Response
A single integer value, or a string.
Example
:SENSe3:MEASure2:EALarm:POINt? ->0
:SENSe3:MEASure2:EALarm:POINt? DEF ->0
:SENSe3:MEASure2:EALarm:POINt? SET ->0
:SENSe3:MEASure2:EALarm:POINt? LINKED ->1,2,3,4
:SENSe3:MEASure2:EALarm:POINt? INFO ->
0:AUTO
1:MANUAL
Command
:SENSe<n>:MEASure<m>:EALarm:POINt