Bit Error Rate Tester | BERT 1005
Quantifi Photonics Ltd.
Version
2.04
74
Syntax
:SENSe<n>:MEASure<m>:EDIagram:VERTical?[<wsp><MIN|MAX|DEF|SET|LIST|
ALL|LINKED>]
Description
Query the vertical (y) resolution step size of the eye measurement.
Parameters
MIN: Query the minimum configurable value
MAX: Query the maximum configurable value
DEF: Query the default power-on value
SET: Query the current SET value
LIST: Query the list of discrete options for configurable values (if available)
ALL: Query the values specified by MIN, MAX, DEF, SET, and LIST
LINKED: Query the linked channels
By default, the command Query the current SET value.
Response
A single value, or a comma separated array. When ALL is requested, the format is
"<MIN>,<MAX>,<DEF>,<SET>,<LIST>"
Example
:SENSe3:MEASure2:EDIagram:VERTical? DEF -> 14.000000
Command
:SENSe<n>:MEASure<m>:EDIagram:VERTical
Syntax
:SENSe<n>:MEASure<m>:EDIagram:VERTical<wsp><value|MIN|MAX|DEF>
Description
Set the vertical (y) resolution step size of the eye measurement
Parameters
value: The desired new set point
MIN: Set to the minimum value
MAX: Set to the maximum value
DEF: Restores the default value
Response
No response
Example
:SENSe3:MEASure2:EDIagram:VERTical 2
Command
:SENSe<n>:PATTern<m>:EQBoost?
Syntax
:SENSe<n>:PATTern<m>:EQBoost?[<wsp><MIN|MAX|DEF|SET|LIST|ALL|STEP|
UNIT|LINKED>]
Description
Query the ED equalizer boost.
Parameters
MIN: Query the minimum configurable value
MAX: Query the maximum configurable value
DEF: Query the default power-on value
SET: Query the current SET value
LIST: Query the list of discrete options for configurable values (if available)
ALL: Query the values specified by MIN, MAX, DEF, SET, and LIST
STEP: Query the minimum resolution of the equalizer boost value
UNIT: Query the default unit for the equalizer boost value
LINKED: Query the linked channels
By default, this command Query the current SET value.
Response
A single value, or a comma separated array. When ALL is requested, the format is
"<MIN>,<MAX>,<DEF>,<SET>,<LIST>".
Example
:SENSe3:PATTern2:EQBoost? ALL -> 0.0,21.0,0.0,0.0,NAN
Command
:SENSe<n>:PATTern<m>:EQBoost
Syntax
:SENSe<n>:PATTern<m>:EQBoost<wsp><value|MIN|MAX|DEF>
Description
Set the ED equalizer boost.
Parameters
value: The desired new set point
MIN: Set to the minimum value
MAX: Set to the maximum value
DEF: Restores the default value
Response
No response
Example
:SENSe3:PATTern2:EQBoost 2.4
Command