Bit Error Rate Tester | BERT 1005
Quantifi Photonics Ltd.
Version
2.04
35
7.5.4
Setting the USB IP address for MATRIQ instruments
When connected via USB, the default IP address is 192.168.101.201. This is a static address set during
instrument calibration. If necessary, this address can be changed. Typing the default IP address in a
supported web browser will open the CohesionUI page for the instrument. The Network Interface
configuration controls are available in the SETTINGS page.
The value in the 3
rd
octet of the IP address can be changed to any available value. It is important to
make sure that any other instruments connected to the computer do not share this new IP address, as
there will be an addressing conflict.
Clicking APPLY will write the new IP address to the instrument settings. Once set, the new IP address will
be displayed on the LCD screen on the back of the instrument.
7.5.5
Setting the Ethernet IP address for MATRIQ instruments
The default Ethernet IP addressing method is dynamic, as the DHCP will automatically assign the
instrument an IP address. This address can be found on the back of the instrument on the LCD screen.
While connected over USB, typing in the assigned IP address in a supported web browser will open the
CohesionUI page for the instrument.