©  Q I-201003 - BioShake iQ  - Short Manual - rev005

STEP 1

STEP 2

STEP 3

STEP 4

STEP 5

STEP 6

STEP 7

Program 

P1

 (or 

P2

)

activated

START

STOP

deactivate

   Filling volume

96-well

(standard)

384-well

(standard)

384-well

(small-vol.)

1536-well

(standard)

      10 %

1800 - 2200 rpm

2200 - 2600 rpm

2800 - 3000

2800 - 3000

      25 %

1600 - 2000 rpm

2000 - 2400 rpm

2400 - 3000

2600 - 3000

      50 %

1400 - 1800 rpm

1800 - 2200 rpm

2200 - 2600

2400 - 2600

      75 %

1200 - 1600 rpm

1600 - 2000 rpm

2000 - 2400

2200 - 2600

   Filling volume

0.2 ml tubes

0.5 ml tubes

1.5 ml tubes

2.0 ml tubes

      25 %

1400 - 1800 rpm

1200 - 1600 rpm

1000 - 1300

1000 - 1300

      50 %

1200 - 1500 rpm

1100 - 1300 rpm

1000 - 1200

  900 - 1200

      75 %

1000 - 1300 rpm

1000 - 1200 rpm

  900 - 1100

  900 - 1100

To Start press 

P1

(or

P2

)

and hold for at 
least 2 sec.

SAFETY

RECOMMENDED MIXING SPEEDS

For MICROPLATES   depending on filling volume/well [%] for diluted fluids

For TUBES depending on filling volume/ well [%] for diluted fluids

Short press 

P1

(or

P2)

PROGRAMMING P1 and P2

Delete program

Yes /  No ? 

e

g

n

a

h

C

           values 

             with the

            cursors!

 Time

Change values with the 
cursors as for progr. step I

 Time         Mix         Temp.

Short press 

P1

(or

P2)

to go forward 
to the next 
program step

Repeat STEP 4-5 
for program 
step II and
step III

To End press 

P1

(or

P2

)

and hold for at 
least 2 sec.

TO RUN THE PROGRAM

Short press 

P1

(or

P2)

Short press

Short press

or

If you do not press Start/
Stop, the program will stop,           
when it is finished.

Short press 

P1

(or

P2)

IMPO RTANT NOTES:

Do not operate the device in rooms where work is being carried out with explosive substances.

Do not use this device to process any explosive, radioactive or highly reactive substances.

Do not use this device to process any substances, which could create an explosive atmosphere.

Danger! Electric shock from damage to device/ power cable. Danger! Lethal voltages inside the device.

Warning! General hazard point.

Warning! Injury from rapidly rotating holder. Injury from rapidly rotating imbalance compensation. Injury from flying tubes and plates.

Warning! Injury from sample material being flung out. Injury from incorrect vortexing.

Warning! Damage due to incorrect power supply.

Caution! Damage to the display from mechanical pressure. Severe vibration. 

Caution! Damage to electronic components caused by spilled liquids. Damage to electronic components from condensation.

Caution! Poor safety due to missing operating manual. Caution when using aggressive chemicals.

Please start even with minimal mixing frequencies to avoid overloading.

O nly mix in sealed tubes and plates. Sample material can be flung out of open, inadequately sealed or unstable tubes and plates.

When working with hazardous, toxic and pathogenic samples, always comply with the nationally specified safety environment.

Pay particular attention to personal safety gear (gloves, clothing, glasses etc., the extraction hood and the safety class of the laboratory.