1. Safety instructions 

1.1 Installation and commissioning 

Place the unit on a plain, dry and clean place.  

Before switching on the unit, leave it standing on its feet for 24 hours to ensure 
that the oils that may have been circulation during transport have settled. 

Pay also attention that the connecting cable  is untangled and free and that nothing is 
resting on it. For direct product cooling use the cooling coil offered by the manufacturer 
only  (optional  item).  It  is  imperative  to  provide  sufficient  ventilation  of  the  unit.  The 
ventilation slots must  remain free for  sufficient evacuation of  heat. No objects may be 
placed under the unit!  
 
The following safety measures must be observed: 
 

  Min. water temperature: -6 °C (with propylene glycol) 

  Max. water temperature: +30 °C 

  Max. ambient temperature: +32 °C 

 

ATTENTION:

 For cold water temperatures below 0 °C it is mandatory, to add 

30 % of propylene glycol to the cooling water !  
Note: There can be no ice bank building when using Glycol.

 

  Keep the unit clean from any dirt, fiber etc. 

  Make sure to connect the unit to the required supply voltage 

  Protect the unit from moisture, no fluids may enter the electric parts 

  Observe the warning and safety information on the electrical components and in 

this instruction manual 

 

1.2 Safety information 

The  unit  should  be  connected  and  operated  by  trained  personnel.  Adjustments, 
maintenance and repairs should be performed by qualified personnel. 
 
Correct function and reliability of operations of this unit can only be warranted, if during 
operation and service the general common saftey precautions,  as well as the specific 
safety instructions described in this manual will be carefully observed.   
 
Handling and treatment of the unit not according to the safety instructions required for 
electrical  appliances  and  the  instruction  in  this  document  may  result  in  severe  bodily 
harm  which  is  not  under  the  responsibility  of  the  manufacturer.  Injury  and  property 
damage can accure through: 

 inappropriate operation  

 incorrect installation or operation 

 improper removal of the necessary protective cover or housing, wetting the electrical 

parts 

 Opening of the unit during operation, which is prohibited 

 

 
 

Summary of Contents for GLY401

Page 1: ...NICHILLY SB COMPATIBLE 115V WTG Quantor GmbH Europa Allee 53 D 54343 F hren Phone 49 0 6502 999 95 0 Fax 49 0 6502 999 95 75 eMail info quantor technology Homepage www quantor technology GB Attention...

Page 2: ...4 1 6 Service 4 2 Intended use 4 3 Commissioning 5 3 1 Connection 5 3 2 Chart 6 4 Decommissioning 7 5 Temperature setting 7 6 Cleaning of the cooling unit 9 7 Troubleshooting 9 8 Noise emission 10 9 T...

Page 3: ...clean from any dirt fiber etc Make sure to connect the unit to the required supply voltage Protect the unit from moisture no fluids may enter the electric parts Observe the warning and safety informa...

Page 4: ...consulted with the manufacturer The unit may only be stored in a dry dust free environment at temperature of 0 C to 40 C 1 5 Electrical installation All electrical installations are to be carried out...

Page 5: ...used to connect the plastic tubes Such kits are not provided with the unit If used please connect them according to the instructions of the manufacturer 4 Close the lid Do not operate the unit when i...

Page 6: ...omp 1 Water level gauge and tank emptying 4 Thermostatic controller 2 Cover fastener 5 Small pipes for water return flow 3 Connections 3 8 5 Expansion valve 8 Water tank 6 Condenser 9 Evaporator cooli...

Page 7: ...the unit with clear water only without Glycol or remnants of Glycol in the water system Recommended set point temperature for using an ice bank 1 0 C 30 2 F ATTENTION to avoid damage on the unit free...

Page 8: ...water mixture approx 30 propylene glycol Please pay attention that the water bath temperature must not fall below 6 C and must not exceed 30 C Controller activation for a minimum set point down to 6...

Page 9: ...Starting device at the compressor is defect Interference of the compressor interwinding fault Establish main connection plug the unit Fill in water Switch on the Thermostat Change the Thermostat Exch...

Page 10: ...s steel Mobile on wheels 10 Declaration of conformity We WTG Quantor GmbH D 54343 F hren declare in sole responsibility that the cooling units MiniCHILLY 05 MiniCHILLY 09 and MiniCHILLY 17 which this...

Page 11: ...11...

Page 12: ...E D I E N U N G S A N L E I T U N G I N S T R U C T I O N M A N U A L I N S T R U C T I O N S D E S E R V I C E M A N U A L D E I N S T R U C C I O N E S Doc ID 2018 06 SBComp115V www quantor technol...

Reviews:

Related manuals for GLY401