HAC 4500 Service Note 1201-022 

Quantum Design 

Service Note 1201-022, Rev. A0 

1

 

 

November 2, 2011

 

Routine Outdoor Helium Compressor Maintenance 

Instructions 

Introduction 

Every month of operation the outdoor compressor of the HAC4500 will need to be thoroughly 
cleaned and serviced to ensure many years of reliable operation.  This maintenance is best performed 
by a trained Heat, Ventilation and Air Conditioning (HVAC) technician or building facility 
personnel. 

In practice, the HAC 4500 maintenance should be accomplished in approximately 1.5 hours and will 
not require the PPMS EverCool II system to be warmed up.  This Service note can be used as a 
guideline for an HVAC technician to familiarize himself with the different routine maintenance 
requirements of the HAC 4500 outdoor helium compressor. 

Tools Needed 

o

 

Medium size Philips screwdriver 

o

 

Water Hose 

Parts Needed 

o

 

Spare 8-32 screws (zs-8csm.38tp) 

I.  Scheduling a maintenance event 

In the PPMS EverCool II the most evident symptom that Cold Head maintenance might need to 
be performed is that the system will not be able to maintain a constant liquid helium level. 

II.  Shutting down the system for Service Maintenance 

The Outdoor Compressor unit maintenance should be scheduled every 2500 hours of operation 

or when overheating of the compressor capsule is observed.  The maintenance can be 
accomplished in just about two hours and should not require the PPMS EverCool II system to be 
completely warmed up. 

1.

 

Set the system in standby mode from MultiVu by clicking Instrument 

Æ

 Shutdown 

Æ

 click 

OK. 

2.

 

Power down the indoor compressor by toggling the switch in the back of the indoor unit. 

3.

 

Power down high voltage power by turning off the high voltage breaker.  WAIT FOR 30 

MINUTES BEFORE PROCEEDING TO THE NEXT STEP!!  This will allow all the 
capacitors in the compressor discharge to a safe level before opening the fuse bay. 

Summary of Contents for HAC 4500

Page 1: ...river o Water Hose Parts Needed o Spare 8 32 screws zs 8csm 38tp I Scheduling a maintenance event In the PPMS EverCool II the most evident symptom that Cold Head maintenance might need to be performed is that the system will not be able to maintain a constant liquid helium level II Shutting down the system for Service Maintenance The Outdoor Compressor unit maintenance should be scheduled every 25...

Page 2: ...and 380 408 V power outlets III Removal and cleaning of the Air Filters 1 Remove the four Air Filters on each side of the compressor by sliding each one up and out of the frame Figure 1 Remove the four Air filters from the compressor frame 2 Using a water hose thoroughly wash the filters and brush off any dirt and debris A soft brush can be used during this operation Figure 2 Wash the compressor a...

Page 3: ...ure onto which the coils are resting 3 Note any rust spots which might require additional care and touch ups Figure 3 Wash the Heat Exchanger coils V Cleaning Compressor Interior 1 Remove at least two of the four panels by unscrewing the 16 screws from each panel Be careful not to cross thread the screws as they might have dirt and rust on them If any of the screws are damaged replace them with ne...

Page 4: ...refully directing the water jet wash out any dirt and debris in the lower shelf of the compressor Make sure to keep the jet away from electrical connections as much as possible see figure 6 WARNING Please note that no water should be directed or otherwise enter through the electronics bay vents This could cause malfunctioning of the equipment Figure 6 Wash the bottom shelf of the compressor unit ...

Page 5: ...ning every screw until tight The Routine maintenance is now complete and the system can be re started VII Starting the PPMS EverCool II System back to operation 1 Restart the helium compressor o Turn the High Voltage Power back on by switching on the High Voltage Power Breaker and wait 5 minutes o Power on the indoor compressor unit by toggling the switch on the back of the indoor compressor unit ...

Page 6: ... outdoor unit informs the user that all communications are established with the outdoor unit Depending on the specific version of the Compressor Firmware you are running you might also see one more message Capsule temperature at 999 00 C is a default value it is displayed every time the power is lost again all normal here 3 If everything is communicating at this point verify that the compressor an...

Page 7: ...022 Rev A0 7 November 2 2011 5 Set the temperature to 300 K Instrument Temperature When you set the temperature to 300 K the equipment automatically opens the valves and starts pumping out the lines 6 Wait five minutes for the system to stabilize 7 Set a new temperature and activate an option ...

Reviews: