18
1.
Identify
[F1 ID ?]
What is the ID number of the sample holder being controlled?
......[F1 ID 14]
Sample holder is a
t2
or other single cuvette sample holder.
Assigned Identities:
ID = 00 –
specialty sample holder
(see command class 14)
14 ‐
t2
24 ‐
t2x2
34 – reserved for turret or linear multi‐sample holder
2.
Controller Firmware Version
[F1 VN ?]
What is the version number of the controller firmware?
......[F1 VN 1.00]
The controller is operating firmware version number 1.00.
3.
Stirrer
[F1 MS ?]
What is the maximum stirrer speed?
......[F1 MS 2500]
The maximum stirrer speed is 2500 rpm.
[F1 LS ?]
What is the lowest stirrer speed?
......[F1 MS 300]
The lowest stirrer speed is 300 rpm.
[F1 SS S 1000]
Set stirrer speed to 1000 rpm.
[F1 SS S 0]
Turn stirrer off and set the speed to zero.
[F1 SS +]
Turn stirrer on and set it to the most recent non‐zero stirrer speed setting.
[F1 SS -]
Turn stirrer off.
[F1 SS ?]
What is the current stirrer speed setting?
......[F1 SS 1000]
Stirrer speed setting is 1000 rpm.
4.
Temperature Control
[F1 TC +]
Turn temperature control on.
[F1 TC -]
Turn temperature control off.
5.
Target Temperature
[F1 TT S 23.10]
Set target temperature to 23.10 ºC.
[F1 TT ?]
What is the current target temperature?
......[F1 TT 71.32]
Target temperature is 71.32 ºC.
[F1 TT +]
Turn on automatic reporting of manual changes to the target temperature.